User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Iotech 1200:Data Acquisition System DATE: 04-Aug-97 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 3 NUMBER OF LINES: 97 CONFIGURATION: Fluke 5700A ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R N P M F W 1.002 ASK+ X 1.003 HEAD INITIAL CONDITIONS 1.004 DISP With No power applied to the UUT, remove any installed 1.004 DISP card. 1.005 DISP Remove both top and bottom covers. 1.006 DISP Remove the four screws that secure's the scanning 1.006 DISP module cover to the PCB. Also remove the two screws 1.006 DISP that secure's the module cover to the back panel. 1.007 DISP Reinsert the scanning card into it's edge connector. 1.008 DISP Remove shorting connectors from JP2 and JP3, located 1.008 DISP to the right of the edge connector. 1.009 DISP Connect the IEEE cable from PORT 1 to the UUT. 1.010 DISP Set switch 9 on the IEEE bus address switch to 1.010 DISP the UP position. 1.011 DISP Apply power to the UUT and allow 1 hour of warm up time. 1.012 IEEE *RX[D6000] 1.013 IEEE *CX[D500] 1.014 HEAD {} 1.015 HEAD {* Chassis Offset Calibration *} 1.016 JMP 2.001 1.017 EVAL 2.001 IEEE K12345X 2.002 IEEE E?X[I$] 2.003 IEEE U2X[I] 2.004 MATH mem1 = mem - 128 2.005 JMPF 2.020 2.006 JMPT 2.020 2.007 DISP Connect the Fluke 5700 HI/LO OUTPUT to UUT JP2 HI 2.007 DISP and JP3 LO. 2.008 5700 0V S 2W 2.009 HEAD Please Wait Performing Offset Calibration 2.010 IEEE H0X[D32000] 2.011 IEEE E?X[I$] 2.012 HEAD * Chassis Offset Calibration * 2.013 IEEE U2X[I] 2.014 MATH mem1 = mem - 128 2.015 JMPF 2.020 2.016 JMPT 2.020 2.017 RSLT =Chassis Offset Calibration sucessfully performed 2.018 RSLT =as per paragraph 7.2.1.2 in the Multiscan 1200 manual. 2.019 JMP 2.023 2.020 DISP [MEM1] 2.020 DISP Calibration error. See page 6.72 in manual. 2.021 IEEE EX 2.022 JMP 3.030 2.023 HEAD {} 2.024 HEAD {* Chassis Gain Calibration *} 2.025 HEAD 2.026 JMP 3.001 2.027 EVAL 3.001 5700 -4.4V S 2W 3.002 HEAD Please Wait Performing - Gain Calibration 3.003 IEEE G0,91X[D32000] 3.004 IEEE E?X[I$] 3.005 IEEE U2X[I] 3.006 MATH mem1 = mem - 128 3.007 JMPF 3.013 3.008 JMPT 3.013 3.009 RSLT =Negative Chassis Gain Calibration sucessfully performed 3.010 RSLT =as per paragraph 7.2.1.3 in the Multiscan 1200 manual. 3.011 IEEE E?X[I$] 3.012 JMP 3.014 3.013 DISP [MEM1] 3.013 DISP Calibration error. See page 6.72 in manual. 3.014 5700 4.4V S 2W 3.015 HEAD Please Wait Performing + Gain Calibration 3.016 IEEE G0,91X[D32000] 3.017 IEEE E?X[I$] 3.018 IEEE U2X[I] 3.019 MATH mem1 = mem - 128 3.020 JMPF 3.028 3.021 JMPT 3.028 3.022 HEAD {} 3.023 RSLT =Positive Chassis Gain Calibration sucessfully performed 3.024 RSLT =as per paragraph 7.2.1.3 in the Multiscan 1200 manual. 3.025 IEEE E?X[I$] 3.026 IEEE EX 3.027 JMP 3.030 3.028 DISP [MEM1] 3.028 DISP Calibration error. See page 6.72 in manual. 3.029 IEEE EX 3.030 DISP Reset switch 9 to it's DOWN position. 3.031 HEAD {} 3.032 END